Configuring a device#

Stream settings#

Set the channels, gain and sample rate before each ba_eeg_manager_start_stream():

const uint8_t electrodes = ba_core_device_features_electrode_count(
    ba_core_device_features_get(ba_eeg_manager_get_device_info(manager)));

for (uint8_t i = 0; i < electrodes; ++i)
{
    ba_eeg_manager_set_channel_enabled(manager, BA_EEG_CHANNEL_ID_ELECTRODE_MEASUREMENT + i, true);
    ba_eeg_manager_set_channel_enabled(manager, BA_EEG_CHANNEL_ID_ELECTRODE_CONTACT + i, true);
    ba_eeg_manager_set_channel_gain(manager, BA_EEG_CHANNEL_ID_ELECTRODE_MEASUREMENT + i, BA_GAIN_MODE_X8);
}
ba_eeg_manager_set_channel_enabled(manager, BA_EEG_CHANNEL_ID_SAMPLE_NUMBER, true);
ba_eeg_manager_set_data_stream_rate(manager, BA_DATA_STREAM_RATE_500_HZ);

ba_error status = ba_eeg_manager_start_stream(manager, NULL, NULL);

When nothing is set, a stream measures every electrode at gain 8 and 250 Hz. ba_eeg_manager_load_config() sends the settings without starting, to check them before the start.

Gain

BA_GAIN_MODE_X1, X2, X4, X6, X8 (default) or X12 per electrode.

Bias electrode

ba_eeg_manager_set_channel_bias_feedback(manager, BA_EEG_CHANNEL_ID_ELECTRODE_MEASUREMENT + i) uses electrode i for bias feedback.

Sample rate

ba_eeg_manager_set_data_stream_rate() accepts the rates of the connected device (see BrainAccess C API) and returns BA_ERROR_UNSUPPORTED_DEVICE for others. ba_eeg_manager_read_sample_frequency() reports the rate in use.

Impedance measurement#

Check electrode contact quality before a recording. ba_eeg_manager_set_impedance_mode() switches the stream to impedance measurement: each BA_EEG_CHANNEL_ID_ELECTRODE_MEASUREMENT channel then carries its electrode’s impedance in kΩ. Configure at least one bias electrode.

Mode

Test signal

BA_IMPEDANCE_MEASUREMENT_MODE_OFF

none: normal EEG

BA_IMPEDANCE_MEASUREMENT_MODE_HZ_7_8

7.8 Hz

BA_IMPEDANCE_MEASUREMENT_MODE_HZ_31_2

31.2 Hz

BA_IMPEDANCE_MEASUREMENT_MODE_DR_DIV4

a quarter of the sample rate

Test signal#

ba_eeg_manager_set_output_mode(manager, BA_OUTPUT_MODE_SQUARE_WAVE) replaces the electrode readings with a built-in square wave, to check the signal path end to end. BA_OUTPUT_MODE_NORMAL switches back.

Battery and charging#

ba_eeg_manager_read_battery_info() returns the battery level and whether the device is charging; the battery callback (ba_eeg_manager_set_callback_battery()) reports every change.

Charging settings choose whether the device can be used while it charges and after how many idle minutes it sleeps:

ba_charging_settings settings = {.enabled_on_while_charging = false, .sleep_timeout = 15};
ba_eeg_manager_set_charging_settings(manager, settings);

Firmware update#

ba_eeg_manager_start_update() sends the firmware image named by sBACORE_Config_t::update_path to the connected device and returns BA_ERROR_OK once the device has accepted it and restarts.