Streaming data#

Channels#

A stream carries the channels enabled with ba_eeg_manager_set_channel_enabled() before ba_eeg_manager_start_stream(). Per-electrode and per-axis channels add an index to a base ID.

Channel ID

Type

Value

BA_EEG_CHANNEL_ID_SAMPLE_NUMBER

size_t

Sample number, from 0 at the stream start

BA_EEG_CHANNEL_ID_ELECTRODE_MEASUREMENT + i

double

Electrode i in µV; its impedance in kΩ in impedance mode

BA_EEG_CHANNEL_ID_ELECTRODE_CONTACT + i

bool

Whether electrode i touches the skin

BA_EEG_CHANNEL_ID_DIGITAL_INPUT

uint8_t

Digital (trigger) input level, 0 or 1

BA_EEG_CHANNEL_ID_ACCELEROMETER + 0..2

float

Acceleration along x, y, z in m/s²

BA_EEG_CHANNEL_ID_GYROSCOPE + 0..2

float

Angular velocity around x, y, z in degrees per second

BA_EEG_CHANNEL_ID_STREAMING

uint8_t

1 for a received sample, 0 for a filled-in lost one

Electrode indices run from 0 to the electrode count minus one (ba_core_device_features_electrode_count()). Enable the motion channels only on devices that have the sensors.

Chunks#

The chunk callback receives samples in blocks of sBACORE_Config_t::chunk_size (25 by default):

typedef void (*ba_callback_chunk)(const void* const* data, size_t data_size, void* user_data);

data holds one row per enabled channel, each with data_size values of the channel’s type. Find a channel’s row with ba_eeg_manager_get_channel_index():

static void on_chunk(const void* const* data, size_t size, void* user_data)
{
    ba_eeg_manager* manager = user_data;
    const size_t* sample = data[ba_eeg_manager_get_channel_index(manager, BA_EEG_CHANNEL_ID_SAMPLE_NUMBER)];
    const double* electrode0 = data[ba_eeg_manager_get_channel_index(manager, BA_EEG_CHANNEL_ID_ELECTRODE_MEASUREMENT + 0)];

    for (size_t i = 0; i < size; ++i)
    {
        printf("%zu %f\n", sample[i], electrode0[i]);
    }
}

The arrays are valid only while the callback runs: copy what you keep. Always use data_size: the last chunk of a stream, and chunks in impedance mode, can be shorter.

Lost samples#

Samples lost on the radio link are filled in, so sample numbers stay continuous. A filled-in row has BA_EEG_CHANNEL_ID_STREAMING set to 0 and zeros in its measurement channels; count those rows to measure the loss.

Units#

Electrode values are in microvolts, already scaled by the gain set with ba_eeg_manager_set_channel_gain().

Annotations#

Mark events in a recording with ba_eeg_manager_annotate(). The library stores your text with the number of the sample the device was recording at that moment, timed with the device clock. Read them with ba_eeg_manager_get_annotations() and clear them with ba_eeg_manager_clear_annotations().