Connect#
The Connect screen (sidebar 🔌, Ctrl+1) is where you discover and
connect signal sources. A device must be connected here before you
can record or view anything.
Three cards select the source type:
Card |
Source |
|---|---|
BrainAccess |
BrainAccess headsets over Bluetooth. |
Network |
Any Lab Streaming Layer (LSL) stream on your network. |
Synthetic Data |
Computer-generated test signals — no hardware needed. |
Once something is connected, a Connected Devices section appears above the cards. Every list has a “Filter devices…” search box.
BrainAccess headsets#
Power the headset on and keep it near your computer. The Bluetooth scan runs continuously; discovered devices appear in the list.
Find your device (type in “Filter devices…” if the list is long).
Choose what to launch:
Signal — normal EEG/biosignal streaming.
Impedance — electrode-contact check (see below).
Test signal — the headset’s built-in square wave, not EEG (see below).
Click Connect (or Start Impedance / Start Test Signal).
Notes on the device list:
While a device is connected, scanning pauses — click “Scan for more devices” to look for additional headsets.
A device that stops advertising doesn’t vanish immediately: its row dims with a “Last seen N s ago” caption. It usually means the device is powering off, out of range, or low on battery. It is removed after about 90 seconds of silence.
“No nearby BrainAccess devices detected yet.” — power the device on, bring it closer, then “Scan for more devices”.
Impedance mode#
Impedance streams electrode contact quality instead of signal — use it while seating the electrodes. When contact looks good, disconnect and reconnect in Signal mode to get actual EEG.
Test signal mode#
Test signal makes the headset stream its built-in square wave instead of electrode data. Use it to verify the signal chain; it is not EEG. The View screen shows the square wave raw: the filter, the notch and contact gating are off and cannot be switched on. Your filter settings come back for a headset connected in Signal mode. Reconnect in Signal mode for real recordings.
If a connection fails#
The device row shows the reason and a Retry button:
Message |
What it means |
|---|---|
Connection timed out — try again |
Device out of range or the Bluetooth adapter was busy. |
Could not connect — check the device is nearby and charged |
The device didn’t respond. |
Device is already connected |
The device is in use — possibly by another application or Board instance. |
Lost connection to the backend during connect — try again |
A temporary internal hiccup; retry. |
Network (LSL) streams#
The Network card lists LSL streams found on your network that are not yet connected, shown as “{name} ({type} – {N}Ch, {rate} Hz)”. Click Connect on a stream to add it as a source.
A BrainAccess headset shared as a remote LSL stream is recognised as such: it connects with its headset model name and shows its battery level, just like a locally connected device.
If the list is empty (“No un-connected LSL streams found…”), make sure the application producing the stream is running and on the same network.
Synthetic data#
The Synthetic Data card lists signal-generator templates, shown as “{name} ({N} ch)”. Click Connect to start one — useful for trying out Board without hardware. Synthetic signals are test data, not real biosignals. You can customise generators via templates (see Device templates).
Connected Devices#
Each connected source shows its status and two controls:
Disconnect — stop the source.
👁 — open this device in its own viewer window (handy with multiple devices).
A source that loses its link — the device powered off, moved out of range, or its stream ended — is removed from Connected Devices automatically within a few seconds. There is no persistent “Offline” row; Disconnect removes a device immediately. Power the device back on and reconnect from its card.
A chip after the device name — ⚠ Metadata or a muted Metadata — means the device’s description is incomplete (for example an LSL stream from another application without a reference electrode). Hover it to see what is missing; see Working with LSL.
The gear icon on a device card opens its template — the channel layout and per-channel settings used when connecting (see Device templates).
An orange FW ↑ badge on a BrainAccess card means newer firmware is available for that headset; see Firmware updates.