Signal quality#

Signal quality scores every electrode while you stream, so you can see what is wrong with a channel and read the specific fix. Open it with the Quality chip in the viewer, the Quality button on the device card (see Connect a headset), or by tapping the N/M good chip.

Note

Opening signal quality does not interrupt the stream and does not change any mode — unlike the Contact check check.

Each electrode gets a quality index from 0 to 1 — lower is cleaner — recomputed about once a second from a four-second window of raw signal. The index and its bands are the same as in BrainAccess Board.

Note

Quality is always computed from the raw signal, so display filter settings cannot change it. Electrode contact is the one thing that does: a disconnected channel is scored unusable, drawn flat, and its PTP reads 0 µV.

Note

A channel whose 0.5–40 Hz signal stays below about 2 µV RMS is scored as flat (index 1, Poor): a stuck electrode, or one that picks up only mains or radio noise and no EEG.

Reading the bands#

Band

Range

Table label

Meaning

Green

< 0.55

Good

Clean — recordable

Amber

< 0.85

Check

Usable but noisy, or still settling

Red

≥ 0.85

Poor

Fix before recording

  • Pill at the top: n/N good.

  • Head map: same 10-20 layout as the Contact check check, coloured by band. Tap any electrode for details.

  • Table: CH and SIGNAL headers are sort buttons (worst-first by default). Tap a row or its chevron for details.

  • How to read signal quality opens the bands and common fixes.

  • The status line warns if the score had to be computed from less than the full gap-free window.

Per-channel details#

Tapping a channel opens a sheet with its score, a diagnosis, a specific fix, and the underlying numbers:

Row

Meaning

Contact

Connected / Not connected, as reported by the headset

DC offset

Electrode offset voltage in µV

Line noise (50/60 Hz)

RMS noise amplitude in the mains band, in µV

Broadband (0.5–40 Hz)

RMS noise amplitude across the EEG band, in µV

Amplitude (p-p)

Peak-to-peak amplitude of the displayed window

Diagnoses and their fixes:

Diagnosis

Fix shown

No contact

Re-seat the electrode: press it firmly to the skin and part the hair underneath so it reaches the scalp

No usable signal (railed or flat)

The channel is railed or saturated: re-seat the electrode and check that its lead is not pulled or shorted

Mains interference

50/60 Hz pickup. Check the reference/ground electrode and move away from chargers and power cables

Movement or broadband noise

Stay still and make sure the electrode is snug — loose contact and motion show up here

Electrode offset or drift

Let it settle for a few seconds, or lift and re-seat the electrode firmly against the skin

Measuring…

Hold still while the signal settles

Good signal

No action needed