Signal quality#
Signal quality scores every electrode while you stream, so you can see what is wrong with a channel and read the specific fix. Open it with the Quality chip in the viewer, the Quality button on the device card (see Connect a headset), or by tapping the N/M good chip.
Note
Opening signal quality does not interrupt the stream and does not change any mode — unlike the Contact check check.
Each electrode gets a quality index from 0 to 1 — lower is cleaner — recomputed about once a second from a four-second window of raw signal. The index and its bands are the same as in BrainAccess Board.
Note
Quality is always computed from the raw signal, so display filter settings
cannot change it. Electrode contact is the one thing that does: a disconnected
channel is scored unusable, drawn flat, and its PTP reads 0 µV.
Note
A channel whose 0.5–40 Hz signal stays below about 2 µV RMS is scored as flat (index 1, Poor): a stuck electrode, or one that picks up only mains or radio noise and no EEG.
Reading the bands#
Band |
Range |
Table label |
Meaning |
|---|---|---|---|
Green |
< 0.55 |
Good |
Clean — recordable |
Amber |
< 0.85 |
Check |
Usable but noisy, or still settling |
Red |
≥ 0.85 |
Poor |
Fix before recording |
Pill at the top: n/N good.
Head map: same 10-20 layout as the Contact check check, coloured by band. Tap any electrode for details.
Table: CH and SIGNAL headers are sort buttons (worst-first by default). Tap a row or its chevron for details.
How to read signal quality opens the bands and common fixes.
The status line warns if the score had to be computed from less than the full gap-free window.
Per-channel details#
Tapping a channel opens a sheet with its score, a diagnosis, a specific fix, and the underlying numbers:
Row |
Meaning |
|---|---|
Contact |
Connected / Not connected, as reported by the headset |
DC offset |
Electrode offset voltage in µV |
Line noise (50/60 Hz) |
RMS noise amplitude in the mains band, in µV |
Broadband (0.5–40 Hz) |
RMS noise amplitude across the EEG band, in µV |
Amplitude (p-p) |
Peak-to-peak amplitude of the displayed window |
Diagnoses and their fixes:
Diagnosis |
Fix shown |
|---|---|
No contact |
Re-seat the electrode: press it firmly to the skin and part the hair underneath so it reaches the scalp |
No usable signal (railed or flat) |
The channel is railed or saturated: re-seat the electrode and check that its lead is not pulled or shorted |
Mains interference |
50/60 Hz pickup. Check the reference/ground electrode and move away from chargers and power cables |
Movement or broadband noise |
Stay still and make sure the electrode is snug — loose contact and motion show up here |
Electrode offset or drift |
Let it settle for a few seconds, or lift and re-seat the electrode firmly against the skin |
Measuring… |
Hold still while the signal settles |
Good signal |
No action needed |